Re: [IPk] Further 508 Experiments

> Dear All,
> Changing my set yesterday afternoon, experimented again.  Twisting
> the end of the syringe this time had no effect.  Shaking the pump,
> however, caused a 'delay' of 0.3 units.  Then I wondered about
> pushing down on the top of the syringe, and was horrified to see a
> huge jet of insulin gush out the end! 

Shaking the pump while it is attached to a "not" hooked up infusion 
line will flex the infusion line and cause its volume to change. It 
will expel insulin under those circumstances. You can easily 
duplicate this without the pump by filling the tube, holding your 
finger over the luer lock end and then shaking the tube. I proves 
nothing. If you effectively seal both ends of the tube by inserting 
it, then no insulin flows when it is shaken. There is nothing to be 
upset about.


email @ redacted
for HELP or to subscribe/unsubscribe, contact: HELP@insulin-pumpers.org
help SUPPORT Insulin Pumpers http://www.insulin-pumpers.org/donate.shtml