[IP] Air in tubbing after priming

I haven't been following this issue closely. However, Minimed has 
several steps in priming. The first is to manually prime until you see 
insulin coming out of the needle. Then you put the syringe in the pump 
and set the pump to prime 5 units of insulin. Then you insert the 
softset or sil, remove the needle from the infusion set, and then prime 
from .5 to 1 unit of insulin. The first two prime steps are to remove 
air from the tubing. The third is to replace the needle volume in the 
infusion set after you remove the needle. Following these steps I have 
never had air in my tubing.