Re: [IP] New Minimed Pump Info

> >3) Temporary Bolus
> >- Temporary bolus can be either insuling rate/hour or a percent of
> >your current basal rate.
> >
> If MMM is truly coming out with a "percentage" temp basal rate, then
> it must be a complete re-design of their current "stepper/DC motor".
> The only way you can get percentage steps (typically of 10%) in a
> "stepper" pump is by being on a 1.0 u/hr rate to begin with.

No. I don't think a stepper/DC motor change is required. For example If you
currently have a basal setting of 0.7 units/hr the pump delivers 0.1 units
every 514 seconds. To decrease this by 10% the 0.1 unit is supplied every
565 seconds.