>   Yes, just disconnect from pigtail or site (we are now on the Quick Sets)
>insert a new set and reconnect using the old tube.

At least in the case of the Sof-Set(R) there are important additional steps 

After disconnecting from the original pigtail you need to connect the old 
tube to the new set and prime enough (4-8 units) till you see insulin 
dripping from the cannula. Now insert the site, remove the introducer 
needle and prime again enough to fill the space left by the removed need 
(0.5 u in the case of the 9 mm Sof-Set).

In the case of the Quick-Set(R) it is impossible to pre-fill the set this 
way before inserting and removing the needle. MiniMed tells me that is one 
of the reason they will not recommend this procedure for the Quick-Set.